201-825-9500 [email protected]

Xenon Trap Monitor

The Xenon Trap Monitor is the perfect complement to the Pulmonex® II Xenon System to ensure a safe environment.

Get a Quote

Description

The Xenon Trap Monitor continuously monitors trap effluent during a xenon study, meeting compliance requirements in most states. An exhaust hose connects from the Pulmonex II exhaust port to the Xenon Trap Monitor’s intake port, making it the perfect complement to the Pulmonex® II Xenon System to ensure a safe environment.

The Xenon Trap Monitor is simple to operate; all controls are located on the front panel. Counting results are displayed with both audible and visual signals to indicate when the xenon trap exhaust port exceeds the threshold of 99 pCi/mL.The unit can also check background levels and perform a self-test for proper operation using a check source (a 10 µCi Cs-137 Check Source is required to calibrate the monitor).

The Xenon Trap Monitor includes all the hardware necessary to mount the unit directly to the Pulmonex® II Xenon System.

Specifications

  • Dimensions:7″ l x 6″ w x 4″ h (15.2 x 17.8 x 10.2 cm)
  • Lead Shielding: .5″ thick (1.3 cm)
  • Input: 22 mm Hose Adapter
  • Detector: Halogen Quenched GM Tube, 1.1 cm dia, 2 mg/cm2 mica window
  • Voltage: 500 volts regulated
  • Buttons: On/Off, Next/Mute
  • Display: 4 Digit LED, 4 RGB Function LEDs (Self Test, Background, Check Source or Count)
  • Readings: pCi/ml or counts
  • Speaker: Internal, beeps in alarm mode
  • Background Count Time: 1 minute
  • Power: 18 volts, UL approved, external; 115 VAC power adapter
  • Weight: 5.25 lb (11.55 kg)
  • Warranty: One year

Manuals

Manual, Xenon Trap Monitor Operation

Models and Ordering

136-755 Monitor, Xenon Trap
Mounts to Pulmonex® II Xenon System

Related:

101-103 Check Source, Cs-137, 10 µCi
Uncalibrated, 1″ dia x .25″ thick (2.5 x .64 cm)