201-825-9500 getinfo@capintec.com

NEMA SPECT Triple Line Source Phantom ™

Designed in accordance with the recommendations by NEMA to standardize the measurement of reconstructed spatial resolution of SPECT

Get a Quote

Description

* Performance Measurements of Scintillation Cameras, NEMA Standards Publication No. NU 1, National Electrical Manufacturers Association (NEMA), Washington, D.C., 2007

Item Numbers
5250-0179 NEMA SPECT Triple Line Source Phantom™

Get a Quote

Features

  • The NEMA SPECT Triple Line Source Phantom is designed in accordance with the recommendations by the National Electrical manufacturers Association (NEMA) to standardize the measurement of reconstructed spatial resolution of SPECT*
  • Material clear is PMMA

Get a Quote

Applications

  • Acceptance testing with NEMA standard
  • Center-of-rotation error evaluation
  • Evaluation of changes of radius-of-rotation on spatial resolution
  • Quantitative evaluation of reconstruction filters and scatter compensation methods
  • Research

Get a Quote

Measurements

  • Cylinder Outside Diameter: 22.2 cm
  • Cylinder Inside Diameter: 20.2 cm
  • Cylinder Outside Height: 23.8 cm
  • Cylinder Inside Height: 20.0 cm
  • Diameter of Line Sources: ~ 1 mm
  • Spacing of Line Sources: 7.5 cm
  • Useful Height of Line Sources: 18.4 cm

Get a Quote